HantekMaster Publish time 2017-5-2 02:24:59

Hantek DSO5102P self calibration failure

Edited by HantekMaster at 2017-5-2 03:26

Hello there,

I bought a DSO5102P that came with firmware 3.40.0(161227.0) and tried to do a self calibration. I 've got the following error message at step 29/38: Self calibration failed, error: 0x10802

I upgraded the firmware to the latest version 3.40.0(170318.0) and getting the same error message.

HW version: 1010xffff8108

Any idea what the problem could be?

Thanks






amy Publish time 2017-5-2 13:30:35

Sir, please replace the two components in the following picture.
1:UD2-4.5NU-L7(relay)
2:74HC595PW(IC)

HantekMaster Publish time 2017-5-3 01:28:50

Thank you Amy for the fast response.

ltkhanh Publish time 2017-8-24 12:23:17

Hello, I updated my DSO5102P to firmware 3.40.0 (170714.0), now the self calibration show error 0x602. Could you tell me how to fix it?
My hardware version is 10000xffff8103
Thank you very much.

amy Publish time 2017-8-25 14:14:33

Sir, please power on the oscilloscope and wait for 20 minutes until it has adequately warmed up. And do self calibration again.

jelka128 Publish time 2018-9-30 19:26:26

Hello, I have MSO5202D, firmware 3.2.35 (151208.0), my hardware version is 1020x55778344. The self calibration show errorx502 . Could you help me how to fix it?
Thank you very much. Roman

ramazzot2009@ho Publish time 2023-2-28 02:31:49




Good afternoon.
I am unable to update my Oscilloscope Firmware DSO5102P. When I try to update, this message appears on the Display: Firmware Update Failed, error 0xfe An update file was not found on the USB device
I download it in WinRar and unzip it but the Oscilloscope shows that there is no file. I am sending pictures of the Display. How to solve it?
My equipment: Oscilloscope HANTEK DSO5102P
Thank you for your attention and solution to the problem.

Translated by Google Translate

PushyG Publish time 2024-4-17 20:25:11

Hello! I have a used scope. Measure funktion displayed fail Voltages. I want run a self calibration, but it freezed step 11/38. I was updated a latest fw, but it freezed again step 12/38 or 11/38. What is the problem? Please help me! Gabor